How to replace the electrode in the EEG cap

This instruction manual is intended for replacing the faulty electrode in MCScap of the BASE, CLINIC, SLEEP, PROFESSIONAL model. Required tools and materials: – Repair kit (BASE / CLINIC / SLEEP / PROFESSIONAL according to the MCScap model); – Resistance measuring device (multimeter); – Phillips screwdriver PH2; – Diagonal pliers; – A clerical or stripping knife (if necessary). Before application this procedure, you need to make sure that all requirement actions were done according to User Manual. MCSCAP REPAIR KIT Repair electrode MCScap – E/T/NT/NTC/NTH/C/TC (electrode type according to MCScap model) – 2 pcs. Extractor – 1 pc. Cable tie 3*100 mm – 4 pcs. Cable fixator – 5 pcs. Thread2m–1pc. Instruction – 1 pc. Timecodes: 00:25 - STEP 1 Identify the faulty electrode on the MCScap. Ensure that the electrode is defective with a multimeter. The device will show a contact break. 00:35 - STEP 2 For silicone screeds and detachable fixators: remove the cable from the defective electrode from the retainer. Be c